A benchmarking protocol for breath analysis: the peppermint experiment

Mr Ben Henderson, Dr Dorota M Ruszkiewicz, Mr Maxim Wilkinson, Dr Jonathan D Beauchamp, Dr Simona M Cristescu, Dr Stephen J Fowler, Dr Dahlia Salman, Dr Fabio Di Francesco, Dr Gudrun Koppen, Dr Jens Langejuergen, Dr Olaf Holz, Miss Andria Hadjithekli, Mr Sergi Moreno, Dr Michele Pedrotti, Professor Dr Pablo Sinues, Miss Gitte Slingers, Dr Michael Wilde, Dr Tommaso Lomonaco, Miss Delphine Zanella, Professor Renato Zenobi, Mr Joan Francis Focant, Dr Stanislas Grassin-Delyle, Dr Flavio Antonio Antonio Franchina, Miss Michaela Malásková, Dr Pierre-Hugues Stefanuto, Mr Giovanni Pugliese, Professor Dr Chris A Mayhew and Professor C. L. Paul Thomas

A benchmarking protocol for breath analysis the peppermint experiment..png

Abstract: Sampling of volatile organic compounds (VOCs) has shown promise for detection of a range of diseases but results have proved hard to replicate due to a lack of standardisation. In this work we introduce the "Peppermint Initiative". The initiative seeks to disseminate a standardised experiment that allows comparison of breath sampling and data analysis methods. Further, it seeks to share a set of benchmark values for the measurement of volatile organic compounds (VOCs) in breath. Pilot data is presented to illustrate the standardised approach to the interpretation of results obtained from the Peppermint Experiment. This pilot study was conducted to determine the washout profile of peppermint compounds in breath, identify appropriate sampling time points and formalise the data analysis. Five and ten participants were recruited to undertake a standardized intervention by ingesting a peppermint oil capsule that engenders a predictable and controlled change in the VOC profile in exhaled breath. After collecting a pre-ingestion breath sample, five further samples are taken at 2, 4, 6, 8 and 10 h after ingestion. Samples were analysed using ion mobility spectrometry coupled to multi-capillary column and thermal desorption gas chromatography mass spectrometry. A regression analysis of the washout data was used to determine sampling times for the final peppermint protocol and the time for the compound measurement to return to baseline levels was selected as a benchmark value. A measure of the quality of the data generated from a given technique is proposed by comparing data fidelity. This study protocol has been used for all subsequent measurements by the Peppermint Consortium (16 partners from seven countries). So far 1200 breath samples from 200 participants using range of sampling and analytical techniques have been collected. The data from the consortium will be disseminated in subsequent technical notes focussing on results from individual platforms.

Download

Previous
Previous

Real-time breath analysis of exhaled compounds upon peppermint oil ingestion by secondary electrospray ionization-high resolution mass spectrometry: technical aspects

Next
Next

Monitoring peppermint washout in the breath metabolome by secondary electrospray ionization-high resolution mass spectrometry